Model Number G10383
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: phone = (b)(6).H3: device evaluation anticipated but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, prior to use while flushing the device with saline, the side-arm separated from a flexor balkin guiding sheath's hub.Another device was used to complete the procedure.The patient did not require any additional procedures or experience any adverse effects due to this event.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, prior to use while flushing the device with saline, the side-arm separated from a flexor balkin guiding sheath's hub.Another device was used to complete the procedure.The patient did not require any additional procedures or experience any adverse effects due to this event.Investigation evaluation: reviews of the complaint history, device history record (dhr), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted, as well as a supplier investigation.The prior-to-use complaint device was returned to cook for investigation.The side-arm was detached.A small amount of glue residue was noted where the side-arm had been connected.A document-based investigation evaluation was performed.A review of the device history record and complaint history found no relevant discrepancies or additional complaints on the final or sub-assembly lots.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A supplier investigation was conducted for the affected component.The supplier reviewed the dhr and manufacturing specifications and concluded that it was possible that the adhesive on the clear tube was insufficient.The information provided upon review of the dmr, dhr, supplier investigation, and investigation of the returned device suggests that there is evidence the device was manufactured out of specification; however, there is no evidence that additional devices manufactured out-of-specification exist in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that supplier manufacturing and quality control deficiencies contributed to this event.Responsible personnel were notified.The risk analysis for this failure mode was reviewed, and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR BALKIN GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key23598097
Report Number1820334-2025-01435
Device Sequence Number10380184
Product Code DYB
UDI-Device Identifier00827002103832
UDI-Public(01)00827002103832(17)270702(10)16137616
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG10383
Device Catalogue NumberKCFW-6.0-38-40-RB-BLKN
Device Lot Number16137616
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2025
Supplement Date Received by Manufacturer04/14/2026
Initial Report FDA Received Date11/19/2025
Supplement Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-