Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the initial reporter phone is not available / reported.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (0932403s) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported issue documented in the complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and / or device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered later.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by johnson & johnson medtech, or its employees that the report constitutes an admission that the product, johnson & johnson medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during a coil embolization of an internal iliac artery prior to endovascular aortic repair (evar) to treat an abdominal aortic aneurysm, devices were used in accordance with the instructions for use (ifus).After the 4 fr diagnostic catheter and a carry león microcatheter (utm) approached the internal iliac artery, embolization was initiated.The first coil used was a 20mm x 60cm deltafill 18 (dlf182060 / 0932403s), but the sheath split at approximately 20-30cm from the tip during the coil insertion, causing the coil to protrude making insertion difficult.Consequently, the coil was removed.It was replaced with another 20mm x 60cm deltafill 18 (dlf182060 / 0932403s).After two additional coils were placed, the 20mm x 60cm deltafill 18 (dlf182060 / 0932403s) was used as the second coil.It was reported that when the 20mm x 60cm deltafill 18 was inserted into the microcatheter, the coil became ¿trapped at the hub¿ and could not advance.Upon withdrawing the coil from the patient¿s anatomy, the coil and the delivery wire were observed to be deformed.It was then replaced with another 20mm x 60cm deltafill 18 (dlf182060 / 0932403s).Three (3) additional cerenovus coils were placed.A total of seven (7) coils were implanted, and the stent graft insertion was initiated.On 18-nov-2025, additional information was received.Per the information, the coil was still attached to the delivery system when it was removed.
|