Blank fields on this form indicate the information is unknown or unavailable.H3: device evaluation anticipated but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, in the middle of an unknown procedure, a flexor high-flex ansel guiding sheath separated.The procedure was converted to an open surgery to retrieve the broken fragments.It is unknown if all fragments were retrieved or how the procedure was completed.According to the initial reporter, the patient did not experience any adverse effects due to this event.Additional information has been requested.
|
|
Summary of event:
as reported, in the middle of an unknown procedure, a flexor high-flex ansel guiding sheath separated.The procedure was converted to an open surgery to retrieve the broken fragments.It is unknown if all fragments were retrieved or how the procedure was completed.According to the initial reporter, the patient did not experience any adverse effects due to this event.Additional information was requested; however, a response was not received.Investigation evaluation:
reviews of the complaint history, device history record (dhr), drawings, instructions for use (ifu), and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.A review of the device history record and complaint history found no relevant discrepancies or additional complaints on the final or sub-assembly lots.The product ifu states ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal."
a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, and ifu suggests that there is evidence the device was manufactured to specification.Based on the information provided and the results of the investigation, cook has concluded that a component failure, unrelated to design or manufacturing, likely contributed to this event.Attempts to obtain additional information from the customer were unsuccessful.The risk analysis for this failure mode was reviewed, and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|