| Model Number |
1400300-08 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure performed was to treat a de novo, moderately calcified, heavily tortuous right coronary artery that is 99% stenosed.During advancement of the 3.00x8mm nc trek neo balloon dilatation catheter (bdc), resistance between the balloon and lesion were noted.Once at the lesion, the nc trek neo balloon ruptured at 8 atmospheres during the second inflation.The nc trek neo bdc was removed and a non-abbott balloon was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted to the balloon during preparation or during the first inflation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the heavily tortuous and moderately calcified anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during the second inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|