| Model Number |
0285 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing noise.The health care professional (hcp) called technical services (ts) for further review of the presenting electrogram (egm).Upon review, the presenting egm showed there to be oversensing noise on the rv lead.Ts also reviewed the daily threshold and impedance measurements trend and noted there to be a steep rise in the rv lead pacing impedances from 850 ohms to 1183 ohms and a gradual increase in shock impedances from 50 ohms to 110 ohms.The pacing and shock impedances were determined to be within normal limits (wnl).Ts discussed possible causes and provided the rv lead to be extracted.At this time, no adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in low-voltage shock impedance (lvsi) measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high-voltage shock impedance (hvsi) measurements and reduced shock efficacy.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of oversensing was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited oversensing noise.The health care professional (hcp) called technical services (ts) for further review of the presenting electrogram (egm).Upon review, the presenting egm showed there to be oversensing noise on the rv lead.Ts also reviewed the daily threshold and impedance measurements trend and noted there to be a steep rise in the rv lead pacing impedances from 850 ohms to 1183 ohms and a gradual increase in shock impedances from 50 ohms to 110 ohms.The pacing and shock impedances were determined to be within normal limits (wnl).Ts discussed possible causes and provided the rv lead to be extracted.At this time, no adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Search Alerts/Recalls
|