• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD Back to Search Results
Model Number 0285
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited oversensing noise.The health care professional (hcp) called technical services (ts) for further review of the presenting electrogram (egm).Upon review, the presenting egm showed there to be oversensing noise on the rv lead.Ts also reviewed the daily threshold and impedance measurements trend and noted there to be a steep rise in the rv lead pacing impedances from 850 ohms to 1183 ohms and a gradual increase in shock impedances from 50 ohms to 110 ohms.The pacing and shock impedances were determined to be within normal limits (wnl).Ts discussed possible causes and provided the rv lead to be extracted.At this time, no adverse patient effects were reported.This rv lead remains in service.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in low-voltage shock impedance (lvsi) measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high-voltage shock impedance (hvsi) measurements and reduced shock efficacy.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of oversensing was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited oversensing noise.The health care professional (hcp) called technical services (ts) for further review of the presenting electrogram (egm).Upon review, the presenting egm showed there to be oversensing noise on the rv lead.Ts also reviewed the daily threshold and impedance measurements trend and noted there to be a steep rise in the rv lead pacing impedances from 850 ohms to 1183 ohms and a gradual increase in shock impedances from 50 ohms to 110 ohms.The pacing and shock impedances were determined to be within normal limits (wnl).Ts discussed possible causes and provided the rv lead to be extracted.At this time, no adverse patient effects were reported.This rv lead remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23600280
Report Number2124215-2025-84878
Device Sequence Number10291361
Product Code NVY
UDI-Device Identifier00802526431401
UDI-Public(01)00802526431401(17)130624
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/24/2013
Device Model Number0285
Device Catalogue Number0285
Device Lot Number100583
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2025
Supplement Date Received by Manufacturer03/26/2026
Initial Report FDA Received Date11/19/2025
Supplement Report FDA Received Date04/15/2026
Date Device Manufactured06/24/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
-
-