• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR LTD NUCLEUS HYBRID L24; NUCLEUS HYBRID COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LTD NUCLEUS HYBRID L24; NUCLEUS HYBRID COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI24RE (L24)
Medical Device Problem Codes Circuit Failure (1089); Impedance Problem (2950); Therapeutic or Diagnostic Output Failure (3023)
Health Effect - Clinical Codes Undesired Nerve Stimulation (1980); Pain (1994)
Date of Event 10/27/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Per the clinic, the patient experienced pain, facial twitching and performance decrement with the device use.Fluctuating impedances were observed across the electrode array and open circuit was noted on 1 electrode.Reprogramming attempts were made; however, the issue could not be resolved.The device was explanted on (b)(6) 2025, and the patient was re-implanted with another cochlear device during the same surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS HYBRID L24
Common Device Name
NUCLEUS HYBRID COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
MDR Report Key23601655
Report Number6000034-2025-04177
Device Sequence Number21326712
Product Code PGQ
UDI-Device Identifier09321502012089
UDI-Public(01)09321502012089(11)161010(17)181009
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/20/2025
Report Date (Section F) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/09/2018
Device Model NumberCI24RE (L24)
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2025
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/30/2025
Event Location Hospital
Date Report to Manufacturer10/30/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/20/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
Date Report Sent to FDA11/20/2025
-
-