H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.Ifu statements the gore® dryseal flex introducer sheath instructions for use (ifu) was reviewed and the following ifu statements were identified in relation to the reported event.Warnings adequate vessel access is required to introduce the sheath into the vasculature.Careful evaluation of vessel size, anatomy, tortuosity, and disease state (including calcification, plaque, and thrombus) is required to ensure successful sheath introduction and subsequent withdrawal.If vessel is not adequate for access, major bleeding, vessel damage, or serious injury to the patient, including death, may result.¿ if vessel size is smaller than the nominal body od (table 1), major bleeding, vessel damage, or serious injury to the patient, including death, may result.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following information was reported to gore: on (b)(6) 2025, the patient underwent an endovascular treatment for a thoracic aortic aneurysm using gore® dryseal flex introducer sheath.When inserting a 24fr sheath via the patient¿s right access site, resistance was encountered, and angiography of the access vessel revealed vascular injury near the right external iliac artery extending to the right common femoral artery.As a treatment, gore® excluder®aaa endoprosthesis (plc121400j) was implanted at this site.After repairing the vascular injury, when the sheath was reinserted, the edge of the previously implanted leg component became caught, causing proximal migration.Consequently, an additional stent graft was implanted distally.After confirming that the patient¿s vital signs were stable, the sheath was delivered and stent grafts were deployed in the descending aorta, and the procedure was completed.The diameter of the narrowed segment of the right access vessel was reported to be approximately 6.8 mm.The physician mentioned that the access was narrow and showed significant calcification, so the vascular injury was anticipated.
|