| Model Number |
BI70002000 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported regarding alleged inaccuracy and to check the navigation accuracy.No patient was present at the time of event.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3, h6) medtronic representative went to the site to test the imaging system and navigation functions check showed that navigation was accurate on both sides of imaging system, 20cm and 40cm fov's.B01, c19, d14, g02008 are applicable continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: m021083c001, serial/lot #: (b)(6): medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a software analysis was initiated; however, it was determined not to be a software issue.Complaint data analysis found that the event was due to a user workflow issue.Navigation function checks showed that navigation was accurate on both sides of the imaging system at 20 cm and 40 cm fovs.The software was functioning as designed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|