ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS SO 17.0; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign - event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the plastic packaging surrounding the implant was damaged.There was no patient involvement.It was reported that no further information is available.D.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h2, h3, h6, h11 complaint can be confirmed through the returned product analysis.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS SO 17.0
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23603559
Report Number0001825034-2025-03714
Device Sequence Number21292242
Product Code KWA
UDI-Device Identifier00880304514416
UDI-Public(01)00880304514416(17)280927(10)6385060
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberN/A
Device Catalogue Number51-106170
Device Lot Number6385060
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2025
Supplement Date Received by Manufacturer03/16/2026
Initial Report FDA Received Date11/20/2025
Supplement Report FDA Received Date04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1