| Catalog Number |
MC1610 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has not been returned to the manufacturer for evaluation.However, photo was provided for review.The investigation of the reported event is currently underway.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound guided kidney biopsy procedure through normal tissue using the maxcore.During the procedure, the outer cannula of the device failed to fire.No coaxial was used.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided, and it was reviewed.The provided photo shows a maxcore disposable core biopsy instrument which in a half-primed position, with the yellow guard.Based on the photo review, the reported failure to fire event cannot be confirmed.Therefore, the investigation is inconclusive for the reported failure to fire as no objective evidence was provided for review.A definitive root cause for the reported failure to fire issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (method, result, conclusion) section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound guided kidney biopsy procedure through normal tissue using the maxcore.During the procedure, the outer cannula of the device failed to fire.No coaxial was used.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|