| Model Number |
305U225 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Tachycardia (2095); Ventricular Fibrillation (2130); Syncope/Fainting (4411)
|
| Date of Event |
01/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 years and 3 months post implant of this 25mm aortic bioprosthetic valve, a defibrillator was implanted.The reason for the defibrillator was reported as syncope with inducible ventricular tachycardia (vt) and ventricular fibrillation (vf).No additional adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|