• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL LAMITRODE S-8 LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL LAMITRODE S-8 LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3286
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 10/30/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced ineffective therapy.Diagnostics revealed high impedances on both leads.As a result, surgical intervention may be undertaken to address the issue.
 
Additional Manufacturer Narrative
Date of event is estimated.
 
Additional Manufacturer Narrative
Date of event and patient's weight is estimated.A patient experienced high impedances/ ineffective therapy was reported to abbott.As a result, the patient¿s leads were explanted and replaced.Therapy was restored.The device was not returned for analysis; however, a diagnostics report was received and confirms multiple contacts with high impedances.Based on the information received, the cause of the high impedances could not be conclusively determined.
 
Event or Problem Description
Additional information indicates surgical intervention was undertaken on 09dec2025 wherein both leads were explanted and replaced to address the issue.Effective therapy was restored post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE S-8 LEAD KIT, 60CM LENGTH
Common Device Name
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
9723098000
MDR Report Key23606087
Report Number1627487-2025-05738
Device Sequence Number13223714
Product Code LGW
UDI-Device Identifier05414734402231
UDI-Public(01)05414734402231(10)4215207(17)150828
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2015
Device Model Number3286
Device Catalogue Number3286
Device Lot Number4215207
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2025
Supplement Date Received by Manufacturer12/09/2025
Initial Report FDA Received Date11/20/2025
Supplement Report FDA Received Date12/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/22/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS IPG
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight91 KG
-
-