| Model Number |
3286 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
10/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced ineffective therapy.Diagnostics revealed high impedances on both leads.As a result, surgical intervention may be undertaken to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event and patient's weight is estimated.A patient experienced high impedances/ ineffective therapy was reported to abbott.As a result, the patient¿s leads were explanted and replaced.Therapy was restored.The device was not returned for analysis; however, a diagnostics report was received and confirms multiple contacts with high impedances.Based on the information received, the cause of the high impedances could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Additional information indicates surgical intervention was undertaken on 09dec2025 wherein both leads were explanted and replaced to address the issue.Effective therapy was restored post operatively.
|
| |
|
Search Alerts/Recalls
|