Model Number DE2110X13B
Medical Device Problem Codes Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an "error in pressure build-up valve for single-hose system" on a trilogy evo, germany device.There was no harm or injury reported.The device was not in patient use.During the evaluation of the trilogy evo, germany device at the third-party service center, a ventilator inoperative error evt_press_sensors_failure was discovered in the device's event log.The device's system board was replaced to address the issue.Additionally, the device's blower, machine flow sensor and one of the in-line filters are contaminated and needed to be replaced.The air foam inlet filter, muffler and particulate filter will be replaced due to the 4-year preventive maintenance procedure.Replacement of the device's system board also addressed evt_out_press_railed_low and evt_mon_press_railed_low errors discovered in the device's event log.The device failed a functional test for aev failure which was resolved by re-seating the tubing.The device passed final testing.This device has been serviced, inspected, tested, and met acceptance criteria in accordance with all of the applicable customer requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO, GERMANY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23606450
Report Number2518422-2025-111959
Device Sequence Number9444387
Product Code CBK
UDI-Device Identifier00606959054110
UDI-Public(01)00606959054110
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDE2110X13B
Device Catalogue NumberDE2110X13B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2025
Initial Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-