| Catalog Number |
122128146 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Joint Laxity (4526)
|
| Date of Event |
11/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
This patient has a srom pinnacle construct implanted sometime ago.The original date of surgery is unknown.The patient was doing well years after the surgery until recently where patient became unstable and dislocated.Several closed reductions were done over the past year.As they became more frequent, the decision was made to revise the patients construct.Doctor explanted the patients pinnacle acetabular cup, liner, and srom femoral hip ball.The srom stem and sleeve were left in situ.Doctor stated that the acetabular cup abduction angle was too vertical which contributed to the dislocation issue.A new acetabular cup and femoral hip ball were implanted providing satisfactory stability.Doi: unknown.Dor: (b)(6) 2025.Affected side: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: d10 (concomitant).Recaptured codes on h6 (health effect - clinical code).
|
| |
|
Search Alerts/Recalls
|