| Model Number |
10000627 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The tap snapped when using.The part was retrieved no issues.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is unavailable for evaluation.The root cause cannot be determined at this time.A supplemental report will be submitted upon receipt of additional information to support this investigation.
|
| |
|
Search Alerts/Recalls
|