The complaint device has been discarded.As a result, no product failure analysis can be conducted, and device malfunction cannot be confirmed.On (b)(6) 2025, a photo evaluation for the device was completed.Photo evaluation summary: upon visual evaluation of the image provided in the complaint, the implant was observed deflated.As the product involved in this complaint was not received, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.Asymmetry may be attributed to one or more of the following: contracture of the fibrous capsule, seroma or hematoma, development of postoperative breast dysplasia, unilateral discrepancy in muscle development, deflation of the implant, incorrect choice of implant shape or size, and surgical technique.Asymmetry is a known complication associated with these devices and is referenced in our current product insert data sheet.Some breast ptosis is a normal component of the mature breast.Ptosis becomes undesirable when the breast parenchyma predominates below the areola, droops considerably below the inframammary fold and the nipple points downward.Augmentation alone, without consideration of the ptosis can produce a less than desirable cosmetic result known as a "rock in sock" deformity or an increased ptotic appearance.Ptosis is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast ptosis and deflation.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|