Model Number 91390
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the screen on their qube compact monitor would freeze and unexpectedly shut down when the monitor was moved.There was no patient or user harm associated with the event.The customer returned their monitor to the equipment service center (esc) for evaluation.An esc technician evaluated the device where they were able to verify the reported issue of the device shutting down, but they were unable to verify the reported freezing display.Further evaluation of the device showed that the cause of the reported power issue was due to a faulty cpu pcba.After replacement of the cpu pcba, proper device operation was observed and the device was returned to the customer.The cause of the reported power issue was due to a faulty cpu pcba.The cause of the reported screen freezing issue could not be determined, as the issue could not be duplicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS HEALTHCARE QUBE COMPACT MONITOR
Common Device Name
QUBE BEDSIDE
Manufacturer (Section D)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer (Section G)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer Contact
whitney snyder
35301 se center st
snoqualmie, WA 98065
MDR Report Key23609585
Report Number3010157426-2025-00222
Device Sequence Number9249495
Product Code MHX
UDI-Device Identifier10841522100550
UDI-Public(01)10841522100550(11)210401
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K120616
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number91390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2025
Initial Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-