• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SURGICAL RELIATACK; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SURGICAL RELIATACK; STAPLE, IMPLANTABLE Back to Search Results
Model Number RELTACK3X10
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Product Quality Problem (1506); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a laparoscopic abdominal wall incisional hernia repair, after 50 shots with the handle, when an at tempt was made to attach the included reload, it could not be fixed and could not be loaded.Additionally, the orientation of the pin was not vertical.Another handle was then used with the reload to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Correction: h6 (fdd a040609 removed, a0201 added).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9 (return date), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device was difficult to load and the orientation of the pin was not vertical.The reported issues were not confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIATACK
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23610496
Report Number1219930-2025-05408
Device Sequence Number9444057
Product Code GDW
UDI-Device Identifier10884521202399
UDI-Public(01)10884521202399(17)280430(10)N5G1638Y
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K140609
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRELTACK3X10
Device Catalogue NumberRELTACK3X10
Device Lot NumberN5G1638Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2025
Supplement Date Received by Manufacturer12/03/2025
02/19/2026
Initial Report FDA Received Date11/20/2025
Supplement Report FDA Received Date12/04/2025
02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-