| Model Number |
470001-09 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The pott scissors instrument was analyzed and found the reported event with the instrument could not be replicated nor confirmed.Failure analysis could not verify the customer reported event.The instrument was tested on an in-house system.The instrument passed the recognition and the engagement tests.The instrument was fully functional.No product issue was found.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|