MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-05
Medical Device Problem Code Expulsion (2933)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event or Problem Description
The recipient reportedly experienced device extrusion.The device had extruded through the skin.It is noted that during a biopsy, the implant area was disturbed which is believed to have caused visible damage to the device.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear implant.
 
Additional Manufacturer Narrative
Additional information: section d.9.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Additional Manufacturer Narrative
Additional information section: h.6.Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly healed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical tests and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
ivvet perez
28515 westinghouse place
valencia, CA 91355
MDR Report Key23612267
Report Number3006556115-2025-01553
Device Sequence Number21292302
Product Code MCM
UDI-Device Identifier07630016878008
UDI-Public(01)07630016878008(11)220526(17)250531
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P960058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2025
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date11/20/2025
Supplement Report FDA Received Date12/03/2025
01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
-
-