RESPIRONICS, INC. ALICE NIGHTONE; VENTILATORY EFFORT RECORDER Back to Search Results
Model Number 1133948
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
An alice nightone device was returned to the manufacturer service center with an allegation of unit stops during the night.There was no allegation of serious or permanent harm or injury.No medical intervention was specified by the patient.During the evaluation, customer's complaint was confirmed.During evaluation it was noted that the printed circuit assembly had electrical damage.It was noted that the top enclosure, bottom enclosure, auxilliar cover were damaged battery and door was missing.The device was scrapped.A final report is being submitted.If new information becomes available, a follow up/supplemental report will be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALICE NIGHTONE
Common Device Name
VENTILATORY EFFORT RECORDER
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23612359
Report Number2518422-2025-111987
Device Sequence Number9258559
Product Code MNR
UDI-Device Identifier00606959047341
UDI-Public(01)00606959047341
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K083874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1133948
Device Catalogue Number1133948
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2025
Initial Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-