|
Additional information: section d9 -date returned to manufacturer, device returned to manufacturer, section g3 - date received by manufacturer, section g6 - type of report, section h6 - evaluation codes, section h11 - manufacturer narrative.The cls was returned to saluda for analysis.The device passed visual inspection and functional testing with no anomalies identified.No impedance issues were found during review of the impedance logs.The root cause of the high impedance could not be definitively determined.The evoke scs system surgical guide details potential risks associated with surgery and spinal cord stimulation including, "uncomfortable changes in stimulation (over and/or under stimulation) and the patient may require surgery (including revision, explant, and replacement) as a result.".
|