| Model Number |
H74939171060410 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b: pro code: (product code): fge, lit.G4: premarket / 510(k): k141521, k141597.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery.A 6.0 x 40, 135cm mustang balloon catheter was advanced for dilation.During the first inflation at 18 atmospheres for 30 seconds, the balloon ruptured.The device was removed, and the procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|