• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. OFP FLUSHING PUMP; OLYMPUS FLUSHING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. OFP FLUSHING PUMP; OLYMPUS FLUSHING PUMP Back to Search Results
Model Number 7501349
Medical Device Problem Code Pumping Problem (3016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the flushing pump was doing a continual pumping of fluid (it was not stopping) during inspection for use.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9; g3; h2; h3; h6 and h11.The device was returned to olympus for inspection, and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OFP FLUSHING PUMP
Common Device Name
OLYMPUS FLUSHING PUMP
Manufacturer (Section D)
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
keymed house, stock road
southend on sea, essex SS25Q H
UK  SS25QH
Manufacturer (Section G)
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
keymed house, stock road
southend on sea, essex SS25Q H
UK   SS25QH
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23617332
Report Number9611174-2025-00359
Device Sequence Number15711399
Product Code FEQ
UDI-Device Identifier15019778003528
UDI-Public15019778003528
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K100899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 01/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7501349
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/28/2025
Supplement Date Received by Manufacturer12/08/2025
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date01/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-