• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. DAKO COVERSTAINER; AUTOMATED SLIDE STAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. DAKO COVERSTAINER; AUTOMATED SLIDE STAINER Back to Search Results
Model Number CS100
Medical Device Problem Codes Mechanical Problem (1384); Device Fell (4014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
The brazil customer reported that the dako coverstainer lower door's shock absorbers, which are used to support the door, occasionally descended when it was opened.The agilent representative reported that the shock absorbers were not replaced during the previous preventive maintenance as they were out of stock locally at the time of service.Upon receipt of the shock absorbers in inventory, the field service engineer (fse) performed the service and replaced the shock absorbers on the instrument.The instrument has been repaired; it is fully operational and is available to the user.This issue does not affect the utility of the instrument, and the customer was able to complete staining.User harm was not indicated or reported.While there was no direct, or indirect, user harm reported for this event; it is being reported due to the potential for harm if the event were to reoccur.Therefore, this report is being filed as part of agilent's commitment to due diligence reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAKO COVERSTAINER
Common Device Name
AUTOMATED SLIDE STAINER
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, sgp 76892 3
SN  768923
Manufacturer Contact
benjamin gaither
1834 state highway 71 west
cedar creek, TX 78612
5123328207
MDR Report Key23617647
Report Number3003423869-2025-00049
Device Sequence Number18653258
Product Code KPA
UDI-Device Identifier05700571112571
UDI-Public(01)05700571112571(11)220401
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date (Section B) 11/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCS100
Device Catalogue NumberCS10030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2025
Initial Report FDA Received Date11/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-