ZIMMER BIOMET, INC. A.L.P.S. LOW PROFILE CORTICAL SCREW; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 851235014
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Rash (2033)
Date of Event 10/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: foreign - the event occurred in japan.Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.D10: concomitant medical products, part (lot): 856135010 (67185876), 856135010 (67247858), 856135012 (67247860), 856135012 (67288183), 856206006 (313300), 14-405024 (67270380), 14-405030 (67183928), 14-405032 (66665527), 14-405034 (66882694), 14-405040 (67098674), 14-441283 (65739497), 40626 (749860).The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that a patient had an initial plate and nail system procedure.Three months postoperatively, eczema developed around the surgical site, raising suspicion of metal allergy.Consequently, all implants were removed.The patient is likely a potential carrier of metal allergies.It was reported that no further information is available.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6; h10.The reported cannot be confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A.L.P.S. LOW PROFILE CORTICAL SCREW
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23617982
Report Number0001825034-2025-03742
Device Sequence Number8968689
Product Code HRS
UDI-Device Identifier00887868352393
UDI-Public(01)00887868352393(17)340716(10)66784803
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number851235014
Device Lot Number66784803
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2025
Supplement Date Received by Manufacturer04/27/2026
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-