Model Number H787CT80LPBDVI0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Bacteremia (4846)
Date of Event 12/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
On or about (b)(6) 2023, plaintiff (b)(6) underwent an insertion an angiodynamics smartport port catheter product, model number ct80lpbdvi, lot number 5804580.Upon information and belief, this smartport device consisted of a port body and a polyurethane catheter.The smartport was implanted at (b)(6) for chemotherapy treatment.Plaintiff and his health care providers used the smartport in a normal, customary, intended, and foreseeable manner, namely as a surgically placed device used to make it easier to deliver medications directly into the plaintiff's bloodstream.Moreover, plaintiff's health care providers did not place or maintain the device incorrectly such that it caused the device to "pinch off" or otherwise malfunction.On or about (b)(6) 2023, plaintiff presented to (b)(6) for treatment of a staph bacteremia infection caused by his smartport device, that was confirmed and diagnosed via positive blood cultures, resulting in the removal of the smartport device on or about (b)(6) 2023.These injuries were directly and proximately caused by the wrongful conduct of the defendants and the unreasonably dangerous and defective nature of his smartport, as further alleged herein.
 
Additional Manufacturer Narrative
The customer's reported complaint description of patient infection cannot be confirmed given the patient centric nature of this serious adverse event (sae).No port device was returned for evaluation since there was no report of port device malfunction, damage or performance issue during use.A device history record (dhr) review of the packaging lot revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Sterilization load release records were reviewed for the packaging lot in question; no issues were observed.The port was implanted into the patient more than 1 month prior to the patient's infection.There was no report of port device malfunction or performance issue during use.If the manufacturing/packaging/sterilization of the port device in question was to contribute to an infection in the patient it would have likely occurred shortly after the implant date and not more than 1 month later.Device directions for use (dfu) cautions that each access of the port should be performed using aseptic technique, following the institutions universal precautions.There is no indication from the reported complaint that the manufacturing/packaging/sterilization of the port device could have contributed to the patient's infection.Infection is cautioned in the device dfu as a potential complication of port system use.Labeling review: the direction for use (dfu) provided with the port device contains the following statements: contraindications: presence of infection, bacteremia, or septicemia.Warnings: the device is to be implanted, used, maintained and removed in accordance with current evidence-based guidance for infection prevention from agencies/organizations such as centers of disease control (cdc) or the world health organization.It is recommended that institutional policies and procedures are updated periodically to reflect current evidence-based guidelines for infection prevention.Contamination of the device may lead to injury, illness or death of the patient.Precautions: carefully read and follow all instructions prior to use.After implantation or usage of the port, the system should be flushed to clear infusates and/or blood components in order to maintain patency.If more than one drug is to be administered, between the individual drug applications, flush the system with 5 to 10 ml normal saline for injection to prevent drug interaction.Potential complications: bacteremia, catheter thrombosis, catheter or port erosion through skin/vessel, death, inflammation, infection, implantation site necrosis or scarring of skin over implant area, thromboembolism, thrombophlebitis, tunnel infection, vascular thrombosis.Use and maintenance after implantation or usage of the port, the system should be flushed and locked to clear infusates and/or blood components in order to maintain patency.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint continues to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO VORTEX
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
MDR Report Key23618148
Report Number1317056-2025-00336
Device Sequence Number10037832
Product Code LJT
UDI-Device Identifier15051684016773
UDI-Public15051684016773
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K190559
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberH787CT80LPBDVI0
Device Catalogue NumberCT80LPBDVI
Device Lot Number5804580
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2025
Supplement Date Received by Manufacturer03/18/2026
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date03/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Other;
Patient SexMale
-
-