• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ASPARTATE AMINO TRANSFERASE; NADH OXIDATION/NAD REDUCTION, AST/SGOT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ASPARTATE AMINO TRANSFERASE; NADH OXIDATION/NAD REDUCTION, AST/SGOT Back to Search Results
Catalog Number 05850819190
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable results from the cobas 8000 c702 module.The customer suspected interference, as the sample in question was collected 17 hours after the patient had received red blood cells due to being anemic.The initial ast (aspartate aminotransferase) result was 17.6 u/l with an analyzer alarm.The initial alt (alanine aminotransferase) result was 11.3 u/l with an analyzer alarm.Due to the alarm, a 1:10 auto dilution was performed.The result for ast was 2.2 u/l, and the result for alt was 10.3 u/l.A new sample was collected from the patient on (b)(6) 2025, and the ast and alt results had no flags.No specific data was provided.This medwatch is for ast (aspartate aminotransferase).Refer to the medwatch with a1 patient identifier (b)(6) for the alt (alanine aminotransferase) assay.
 
Additional Manufacturer Narrative
The cobas 8000 c702 module serial number was (b)(6).The investigation is ongoing.
 
Additional Manufacturer Narrative
On (b)(6) 2025, further testing of the sample generated the following results: alt 6.8 u/l with a data flag and 7.8 u/l with a data flag.Ast 16.7 u/l with a data flag and 12.9 u/l.The investigation is ongoing.
 
Additional Manufacturer Narrative
Medwatch fields d4 catalog no and primary udi number, and g4 pma / 510k (premarket numbers) were updated.
 
Additional Manufacturer Narrative
Based on the information provided, the investigation determined that the event was consistent with the sample containing a substance that strongly absorbs at 340 nm, the measuring wavelength of the assay.The event was a sample-specific issue.There was no product problem identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPARTATE AMINO TRANSFERASE
Common Device Name
NADH OXIDATION/NAD REDUCTION, AST/SGOT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key23618537
Report Number1823260-2025-04768
Device Sequence Number9072612
Product Code CIT
UDI-Device Identifier04015630928934
UDI-Public04015630928934
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05850819190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/31/2025
Supplement Date Received by Manufacturer12/19/2025
01/26/2026
04/08/2026
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date01/13/2026
02/18/2026
04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-