• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO2 EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO2 EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4030
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that cautery functionality was compromised and was restored once the defective hemopro2 extension cable was swapped out for a new one.
 
Additional Manufacturer Narrative
(b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Additional Manufacturer Narrative
(b)(4).Updated sections: b4,b5,d4 lot#, exp date, udi#,d9,g3,g6,h2,h3,h6,h11.The device was returned to the factory for evaluation on 12/18/2025.An investigation was conducted on 12/22/2025.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the intact cable.A mechanical evaluation was conducted using a reference adaptor, power supply and hemopro 2 device.The cable was able to be connected to the adaptor with no physical or visual difficulties.The reference harvesting device was attached to the cable with no physical or visual difficulties observed.An electrical evaluation was conducted.A pre-cautery test was performed per the instruction for use (ifu) with a reference hemopro 2 evh tool, reference adapter and reference power supply set at level 3.0.The device passed the pre-cautery test.The power supply emitted an audible beeping sound and the evh reference tool produced steam and heat.An activation and transection capability test was performed over four (4) repetitions using "max life test method.The reference device successfully transected tissue four (4) times.Based on the returned condition of the device and the results of the evaluation, the reported failure "electrical /electronic property problem" was not confirmed.The reported device is an oem device.The certificate of conformance was reviewed for the lot #12411138.The vendor certifies that this device lot # conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
 
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure, the defective hemopro2 extension cable was swapped out for a new one as the cautery functionality was compromised, it was restored once the defective cable was swapped.There was no delay.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO2 EXTENSION CABLE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key23618616
Report Number2242352-2025-0001217
Device Sequence Number15943166
Product Code GEI
UDI-Device Identifier00607567700925
UDI-Public(01)00607567700925
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-4030
Device Catalogue NumberVH-4030
Device Lot Number12411138
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer12/10/2025
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date12/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-