| Model Number |
DDU-100B |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported their battery pack will not stay latched in their aed.The customer did not report this occurring during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Although requested, the aed has not been returned and the cause of the complaint is not known.
|
| |
|
Additional Manufacturer Narrative
|
|
The unit was requested for return and further evaluation.Upon return, the device underwent bench testing, it was confirmed that the battery pack could not be securely latched into the battery well.The issue was caused by broken battery latch, which are intended to assist in securing the battery in place.Despite the damaged latch, the device is able to function when the battery is manually held in position.
|
| |
|