It was reported by medwatch picc line was inserted into patient after selection of basilic vein.Subsequent imaging revealed a metallic density as a probable portion of wire that extended from the region of the left pulmonary artery into the left lower lobe.The patient underwent a right heart cath and the metal was successfully removed in two pieces.A 4 cm fragment was removed from the left pulmonary artery.A 1 cm fragment was removed from the inferior vena cava.The fragments were described as thin, black, soft wire-like material reportedly matching the stylet that accompanies power picc kit.No other information was provided.
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been received by the manufacturer for evaluation.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.
|