MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number FR2110X14B
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a damaged screen was found on a trilogy evo device.The device was not in use on a patient at the time of the occurrence.The trilogy evo device was returned to the third-party service center for evaluation, and the customer¿s complaint was confirmed.During the evaluation, the third-party service technician provided no findings on the cause of the damage display that was on this device.In conclusion, the device's display was replaced to address the issue.Additionally, during the service of the device, the muffler assembly, particulate filter, and air inlet foam filter were replaced for preventative maintenance unrelated to the reportable issue.The vanity cover was replaced due to the alarm audio pause button missing on the device.This was unrelated to the reportable issue.The internal battery door and blower chassis were replaced due to physical damage unrelated to the reportable issue.The machine flow sensor was replaced for contamination unrelated to the reportable issue.The device passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23621907
Report Number2518422-2025-112009
Device Sequence Number9072720
Product Code CBK
UDI-Device Identifier00606959055506
UDI-Public(01)00606959055506
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberFR2110X14B
Device Catalogue NumberFR2110X14B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2025
Initial Report FDA Received Date11/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-