COOK INC ROADRUNNER THE FIRM LT HYDROPHILIC WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Model Number G06866
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: device evaluated by mfg = device evaluation anticipated but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, a hair was discovered inside the packaging of two units of the 'roadrunner the firm lt hydrophilic wire guide'.The issue was noted during device inspection within a distribution facility, therefore, there is no patient involvement.
 
Event or Problem Description
During preliminary device analysis of returned product on (b)(6) 2025, an additional roadrunner the firm lt hydrophilic wire guide was found to have foreign matter inside the packaging, making the total quantity three units.
 
Additional Manufacturer Narrative
Correction: b5 additional information: d9 h3: device evaluated by mfg = device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Summary of event: as reported, a hair was discovered inside the packaging of two units of the 'roadrunner the firm lt hydrophilic wire guide'.The issue was noted during device inspection within a distribution facility, therefore, there is no patient involvement.During preliminary device analysis of returned product on 02dec2025, an additional roadrunner the firm lt hydrophilic wire guide was found to have foreign matter inside the packaging, making the total quantity three units.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint devices was also conducted.The complaint devices were returned to cook for investigation.A hairlike fiber was confirmed within the packages.A document-based investigation evaluation was performed.A review of the device history record found that one device from the complaint lot did not pass quality control inspections; however, the affected product was scrapped prior to release from cook.A review of complaint history found no additional related complaints for the complainant lot.The product ifu states, ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the returned devices, complaint file, ifu, dmr, and dhr suggests that there is evidence that the device was manufactured out of specification.However, cook has concluded that this issue is an isolated incident, and there is no evidence of additional nonconforming devices in house or out in the field.Although one device from the complaint lot did not pass quality control inspections, the product was scrapped prior to release from cook, there are 100% inspections in place, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot-related complaints have been received from the field.Therefore, it was concluded that there is no evidence that devices manufactured out-of-specification exist in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded manufacturing and quality control deficiencies contributed to this incident.The risk analysis for this failure mode was reviewed, and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROADRUNNER THE FIRM LT HYDROPHILIC WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key23621917
Report Number1820334-2025-01449
Device Sequence Number15697524
Product Code DQX
UDI-Device Identifier00827002068667
UDI-Public(01)00827002068667(17)280501(10)16583269
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
K182985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberG06866
Device Catalogue NumberRLPC-35-145
Device Lot Number16583269
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Supplement Date Received by Manufacturer12/02/2025
02/17/2026
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date12/22/2025
03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1