|
We received an allegation about discrepant results for 1 patient sample tested with elecsys estradiol iii assay on a cobas pure e 402 analytical unit.Initial result: 740 pg/ml.1st repeat result: 49.1 pg/ml.2nd repeat result: 48.4 pg/ml.On (b)(6) 2025: the sample was repeated on a delta instrument, resulting in a 3rd repeat result of 57.2 pg/ml.The sample was repeated 10 more times on a different cobas pure analytical unit, resulting in the following repeat results: 46.2 pg/ml, 51.2 pg/ml, 47.9 pg/ml, 46.6 pg/ml, 47.4 pg/ml, 46.8 pg/ml, 47.2 pg/ml, 46.4 pg/ml, 48.3 pg/ml, and 47.2 pg/ml.The result that was considered correct was around 50 pg/ml.
|