• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number DDU-E2300
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Although requested, additional information regarding the complaint has not been provided and the cause of the complaint is not known.
 
Event or Problem Description
A customer reported their aed is flashing red and prompting a "replace battery pack now" warning.They reported this did not occur during patient use.
 
Additional Manufacturer Narrative
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.The cause of the aed not powering on was related to the customers failure to maintain the battery pack.As a follow up with the customer, the proper maintenance of this device was reviewed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT 06405
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key23622170
Report Number3003521780-2025-01943
Device Sequence Number15697667
Product Code MKJ
UDI-Device Identifier10815098020239
UDI-Public(01)10815098020239
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-E2300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-