| Model Number |
DDU-E2300 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.The cause of the aed not powering on was related to the customers failure to maintain the battery pack.As a follow up with the customer, the proper maintenance of this device was reviewed.Once the new battery was installed and a manual self test performed the aed functioned as designed and was able stay in service.
|
| |
|
Event or Problem Description
|
|
An international distributor reported their customer's aed would not power on, but it powered on with a spare battery.The customer did not report this occurring during patient use.
|
| |
|