| Model Number |
2120 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pump had a downstream occlusion seal that was detached during testing.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.Visual inspection was able to verify the reported problem.The downstream occlusion seal was replaced.The device then passed all functional tests.The service history review had no indication that the complaint was related to a service of the device within the review period.
|
| |
|