• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CS-27702-E
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "on (b)(6) 2025, during the central venous catheterization procedure, the guide wire inside the central venous puncture kit was easy to kink, which subsequently led to the failure of the puncture.Additionally, the ars leak during use on the patient, which affected the operation process; the most serious issue was that the guide wire was extremely easy to kink, and this was the key cause of the puncture failure.Additional information received from the customer states "the doctor thinks our swg is easy to kink".Multiple puncture kits were replaced, causing unneccessary lost and injury to the patient.Associated mdr number includes: 3006425876-2025-01100, 3006425876-2025-01099.
 
Event or Problem Description
It was reported that "on (b)(6) 2025, during the central venous catheterization procedure, the guide wire inside the central venous puncture kit was easy to kink, which subsequently led to the failure of the puncture.Additionally, the ars leak during use on the patient, which affected the operation process; the most serious issue was that the guide wire was extremely easy to kink, and this was the key cause of the puncture failure.Additional information received from the customer states "the doctor thinks our swg is easy to kink".Multiple puncture kits were replaced, causing unneccessary lost and injury to the patient.Associated mdr number includes: 3006425876-2025-01100, 3006425876-2025-01099, and 3006425876-2025-01136.
 
Additional Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Common Device Name
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key23623806
Report Number3006425876-2025-01136
Device Sequence Number18400013
Product Code FOZ
UDI-Device Identifier00801902102362
UDI-Public(01)00801902102362(11)241003(17)260912(10)71F24J1433(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K900263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/12/2026
Device Catalogue NumberCS-27702-E
Device Lot Number71F24J1433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2025
Supplement Date Received by Manufacturer12/30/2025
Initial Report FDA Received Date11/21/2025
Supplement Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-