| Catalog Number |
362788 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use of bd vacutainer® ppt¿ plasma preparation tube k2e 9.0 mg, an unspecified number of tubes had a foreign body inside the tube.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes.H.3 device eval by manufacturer? yes.D9: returned to manufacturer on: 09-dec-2025.Investigation summary: bd received samples and photos for investigation, and evaluation of the six samples and seven photos confirmed the presence of fm in the tube.Additionally, 100 retain samples were visually inspected, with no issues being identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter - non-biological.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use of bd vacutainer® ppt¿ plasma preparation tube k2e 9.0 mg, an unspecified number of tubes had a foreign body inside the tube.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|