This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the chuck with key device device was broken and fell apart into two pieces was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device fell apart - unattached.The assignable root cause was determined to be due to component failure from wear.
|
|
It was reported that during service and evaluation, it was determined that the chuck with key device fell apart-unattached and had cosmetic damage.It was further determined that the device failed pretest for general condition and check cannulation of attachment.It was noted in the service order that the device was broken and fell apart into two pieces.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2025.All available information has been disclosed.
|