• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL FLEXIBLE ENDOSCPC CANN 4.5MM; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL FLEXIBLE ENDOSCPC CANN 4.5MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number 72202971
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a femoral tunnel perforation for a knee arthroscopy, the drill endoscopic cannula broke in the middle at the moment the surgeon reversed the drill.One part fell into the cavity and was successfully removed.The procedure was resumed, with a delay of less than 30 minutes, using a similar sn back-up.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL FLEXIBLE ENDOSCPC CANN 4.5MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23630776
Report Number1219602-2025-02964
Device Sequence Number13090595
Product Code NBH
UDI-Device Identifier03596010654526
UDI-Public3596010654526
Combination Product (Y/N)N
Initial Reporter CountryMX
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number72202971
Device Lot Number51067657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/24/2025
Supplement Report FDA Received Date12/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient Age39 YR
Patient SexMale
-
-