MAUDE Adverse Event Report: BECTON DICKINSON BI-EXT SET 15CM MACRO BORE SPIN NUT; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BI-EXT SET 15CM MACRO BORE SPIN NUT; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 385161
Medical Device Problem Codes Leak/Splash (1354); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the connection securement isn¿t functioning as we used to have it before; it is not locking properly leading to unexpected disconnection and dislodgement of the line affecting the patient safety through losing the medication.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
Investigation results: our quality engineer inspected the 2 physical samples submitted for evaluation.The reported issue of connection issues was not confirmed upon inspection of the samples.Analysis of the sample showed that a connection problem was not observed in either of the 2 samples provided.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI-EXT SET 15CM MACRO BORE SPIN NUT
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key23630808
Report Number9610847-2025-00436
Device Sequence Number13106818
Product Code FPA
UDI-Device Identifier00382903851614
UDI-Public(01)00382903851614
Combination Product (Y/N)N
Initial Reporter CountryKU
PMA/510(K) Number
K142527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number385161
Device Lot Number4214810
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2025
Supplement Date Received by Manufacturer01/23/2026
Initial Report FDA Received Date11/24/2025
Supplement Report FDA Received Date01/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-