SYNTHES GMBH CHUCK WITH KEY-REAM SPEED F/BPL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
530.732 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the chuck with key device was frozen/would not move-chuck seized, worn seal, color band was chipping/fading and cosmetic damage-worn coupling.It was further determined that the device failed pretest for markings and check drill chuck function.It was noted in the service order that the device did not work.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from wear.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: correction: device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the device did not work was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device was frozen/would not work-chuck seized.This correction did not change the assignable root cause of component failure from wear.
|
| |
|
Search Alerts/Recalls
|
|
|