Visual inspection: during the investigation, no significant deformations or damage of the valves was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the valve is not operating within the acceptable tolerance in the horizontal position.An accelerated outflow could be derterminde internal inspection: after dismantling of the valve, deposits were found in mininav.Result: in advance, we would like to point out that the product sent in was not submerged in liquid at the time of delivery.The testing of dry valves is only of limited value.The product properties can be influenced by dry residues of cerebrospinal fluid or blood.Despite this, we have examined the valve.Based on our investigation results, we can confirm accelerated drainage and deposits on the mininav.Deposits caused by natural substances found in the body, such as protein, blood, or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic deposits can compromise the integrity of the valve.The examination of the returned product is complete with the preparation of this report.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|