| Model Number |
DHEA-S |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas e 801 analytical unit serial number was (b)(6).The qc was in range prior to the event.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation about discrepant results for one patient sample tested with elecsys dhea-s assay on a cobas e 801 analytical unit.Initial result: 630 ug/dl.The qc went out of range, prompting the repeat of the samples.Repeat result: 417 ug/dl (it is unclear if this result was from the same or a different e 801).The repeat result was deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) found a kinked procell line.The fse replaced and reseated the procell line.Qc was acceptable; however, the instrument check failed.Preventive maintenance was performed.Mechanical checks passed.The instrument was operating within specification.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Search Alerts/Recalls
|