Model Number M00584050
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.
 
Event or Problem Description
It was reported to boston scientific corporation that a dreamtome rx 44 was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat biliary strictures performed on (b)(6) 2025.During the procedure, the wire broke while cutting the papillotomy.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another dreamtome rx 44.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fully recovered.
 
Additional Manufacturer Narrative
Block h2: block d4, h4 have been updated based on the additional information received on november 20, 2025.Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.
 
Event or Problem Description
It was reported to boston scientific corporation that a dreamtome rx 44 was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat biliary strictures performed on (b)(6) 2025.During the procedure, the wire broke while cutting the papillotomy.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another dreamtome rx 44.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fully recovered.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.Block h11: investigation results the returned dreamtome rx 44 was analyzed, and a visual inspection found that the cutting wire was blackened, kinked and broken from the proximal pierce hole.The guidewire was found in good condition.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of cutting wire break was confirmed.The results of the analysis performed on the returned device found that the cutting wire was kinked and broken from the proximal pierce hole.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.The break in the cutting wire could have been generated if there was contact between the device and the scope during energization or if the device exceeded the maximum of voltage during procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wires integrity, causing premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can bend the cutting wire as observed in the product analysis.The investigation findings and all information available concludes the most probable root cause is an adverse event related to procedure.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a dreamtome rx 44 was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat biliary strictures performed on (b)(6) 2025.During the procedure, the wire broke while cutting the papillotomy.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another dreamtome rx 44.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMTOME RX 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key23632942
Report Number3005099803-2025-06308
Device Sequence Number15782227
Product Code KNS
UDI-Device Identifier08714729778271
UDI-Public(01)08714729778271(17)280805(10)37143994
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00584050
Device Catalogue NumberM00584050
Device Lot Number0037143994
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2025
Supplement Date Received by Manufacturer11/20/2025
12/17/2025
Initial Report FDA Received Date11/24/2025
Supplement Report FDA Received Date12/16/2025
01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1