| Catalog Number |
IAB-06840-U |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted into the patient, the balloon had not inflate completely.The balloon was found ruptured after remove the catheter.Change the new catheter".No patient injury or consequence reported.The patients current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The serial number on the returned sample is (b)(6).Returned for investigation was a 40cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) without the original packaging box.The sample was returned in a cardboard box and was in a ziploc bag.Upon return, the teflon sheath was noted on the iabc; blood was noted in the sheath sidearm.The one-way valve was tethered and connected to the short driveline tubing.The bladder was fully unwrapped.The central lumen was noted broken at approximately 1.7cm from the iab distal tip.Dried blood was noted on the exterior surfaces of the returned sample.Some blood was noted within the bladder/helium pathway.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute ac-cording to quality system document.An attempt to aspirate and flush the iabc central lumen using a 60cc lab-inventory syringe was unable to be successfully completed.Upon aspiration, water was unable to be consistently pulled into the syringe.The attempt to flush resulted in bladder inflation, which is consistent with the previously confirmed broken central lumen.The iabc was leak tested in accordance with testing methods from manufacturing procedure.A leak was immediately detected from the iabc distal tip and iabc luer end.The leak from the iabc distal tip and iabc luer end are consistent with the previously confirmed broken central lumen.The iabc was leak tested again with the iabc distal tip and luer end blocked off; no other leaks were detected.A lab inventory 0.025in guidewire was back loaded through iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.Some blood and debris were noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for "the balloon had not inflate completely" is confirmed.During the investigation, the intra-aortic balloon catheter (iabc) central lumen was noted broken and can result in the inability for the bladder to inflate.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the broken central lumen.The root cause of the complaint is undetermined.No further action required at this time.This will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted into the patient, the balloon had not inflate completely.The balloon was found ruptured after remove the catheter.Change the new catheter".No patient injury or consequence reported.The patients current condition is reported as "fine".
|
| |
|
Search Alerts/Recalls
|