|
It was reported during a therapeutic liver resection procedure, the probe of the ultrasonic surgical device broke.The broken piece of the probe was safely removed from the patient.The issue did not affect the patient.A delay of greater than or equal to 30 minutes occurred, however the procedure continued safely and was completed with a back-up device.There were no reports of patient harm.
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d4, d9, h2, h3, h6, h7, h9, h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reported failure was confirmed.The device evaluation found the probe was broken at 15.93 mm from its distal end.The broken probe tip was not returned.A root cause could not be identified.Based on the results of the investigation, a likely mechanism causing the reported event might be the following: 1.The output was activated in seal & cut mode while the grasping section was grasping thick tissue.Therefore, the probe and the tissue pad came into contact at the rear end of the grasping section, causing the tissue pad to wear out.2.The non-insulated area of the grasping section and the probe came into contact due to wear of the tissue pad.3.The output in seal & cut was activated while the non-insulated area of the grasping section was in contact with the probe.As a result, a contact mark developed.4.A force to activate the output in seal &cut mode or a force to grasp tissue was applied to the probe.Therefore, cracks developed at a contact mark.5.A force was applied to the probe causing it to break.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|