| Catalog Number |
0119L20 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter name: (b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that after transurethral bladder surgery, the doctor inserted and performed bladder irrigation, then connected to uroguard.After the anesthesiology procedure, it was confirmed that the foley catheter had been spontaneously removed.Checked for leakage of sterile water.Leakage of sterile water was confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed ¿ manufacturing related.The reported failure was able to be reproduced.The product was used for urological care.Based on the evaluation, a tear was found on the rubberized layer causing interlumen leakage.Hence, this complaint was confirmed related to manufacturing process.A potential root cause for this failure could be insufficient latex strength, low/non-uniform rubberize thickness, incorrect wire pressing technique, incorrect stripping technique etc.A review of the device labelling has been conducted for investigation purposed.A review of the manufacturing process indicates that the process is capable of producing of this type of defect.However, current in-process controls are adequate to identify the defect or verify the product integrity.Corrections: f, h.Initial reporter name: (b)(6).This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that after transurethral bladder surgery, the doctor inserted and performed bladder irrigation, then connected to uroguard.After the anesthesiology procedure, it was confirmed that the foley catheter had been spontaneously removed.Checked for leakage of sterile water.Leakage of sterile water was confirmed.
|
| |
|
Search Alerts/Recalls
|