MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON UNITY CATARACT FMS PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON UNITY CATARACT FMS PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000340
Medical Device Problem Codes Improper Flow or Infusion (2954); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Intraocular Pressure Decreased (4468)
Date of Event 09/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
The physician reported that the cassette displayed some resettable advisories and there was no infusion going into the eye leading to anterior chamber instability during sculpt mode of vitreo-retinal surgery.The surgery was completed after replacing the cassette with another one as well changing changing the equipment.There was patient contact but no impact to the patient.This report pertains to cassette involved in reported events.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The turbid pak was visually inspected, and no obvious defects were found.All manifolds were inserted properly in the cassette housing.The valves on the cassette base were intact.The small part tray was not returned with the product.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The product was tested using the console with the calibrated console.The product could prime and tune with the active sentry handpiece, the 0.9 milli meter aspiration bypass system tip and infusion sleeve from the lab stock successfully.No message code appeared on the screen.The fluidics management system wall leak test, irrigation, aspiration and redundant pressure sensors displayed the correct accuracy readings when evaluated.The maximum vacuum test successfully achieved the target vacuum pressure.The product passed functionality per procedure.The investigation conducted a non-conformance review of the reported lot number.No relevant deviations were identified, all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event as the product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product meeting specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNITY CATARACT FMS PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23633881
Report Number1644019-2025-04989
Device Sequence Number9317462
Product Code LRO
UDI-Device Identifier00380650003409
UDI-Public(01)00380650003409(17)270131(11)250218(10)178WHV
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065000340
Device Lot Number178WHV
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer03/20/2026
Initial Report FDA Received Date11/24/2025
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNITY VCS CONSOLE.
Patient SexUnknown
-
-